Microbial Air Monitoring System
Equip Your Lab for Success
Contact Us for a Wide Range of Analytical Instruments and Equipment!”
TME 003- Microbial Air Monitoring System (AIR SAMPLER)
Intended Use
It is used for monitoring of microbial air contamination in Pharmaceutical and Food manufacturing and clinical environments.
Product Summary and Explanation
Microbial Air Sampler carries a unique structure to avoid recirculation of air and is validated as per BS EN ISO 14698- Part 1, for Physical and Biological efficiency. The upper part includes a sampling head, sampling seat and a pump. The lower part includes controller and batteries. It is of low noise pump sampling, simple operation and stable performance. Sampling port is made of high-quality aluminum, which is suitable for various way of sterilization.
Composition
| SPECIFICATIONS | |
|---|---|
| Nominal Air Flow Rate | 100 litres of air per minute |
| Impact Speed | < 20m/sec |
| Air Sample Volume | 10-9999 liters can be selected and stored in the instrument. Three different volumes can be set |
| Maximum volume of air per cycle | 1000 litres |
| Plate size | 90±5 mm & 55 mm |
| Weight of instrument | 2.68 kg |
| Construction | High quality anodized aluminum |
| Display | High Brightness LCD display |
| Accessories | Empty Petri-plate |
| Battery Pack | Rechargeable Li Ion (without memory effect). Battery Charging takes only 4 hours. After charging, the battery will work up to 8 hours |
Principle
Microbial Air sampler is designed on the theory of isokinetic sampling. Air passes through small holes onto surface of an agar plate with high velocity. At the end of the cycle, the plate is removed and incubated and visible colonies are then counted for an assessment of the level of contamination.